Herunterladen Inhalt Inhalt Diese Seite drucken

Ottobock Genium Gebrauchsanweisung Seite 52

Bionic prosthetic system
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

MyMode Motion Switching
MyMode Motion Switching
Switching from MyModes into basic mode
INFORMATION
Switching back to basic mode by bouncing up and down three or more times cannot be deactivated. Thus, you can
access basic mode even if switching via bouncing is not enabled in X-Soft.
≥ 3 x
bouncing recognised
* Returning to basic mode is also possible by connecting the battery charger
1. 3x bouncing on the forefoot
2. Remove load from prosthesis for 1 second
• Feedback from the prosthesis: 1 long vibration signal
3. Wait for 1 second
• Feedback from the prosthesis: 1 short beep signal
Switching from basic mode to the first MyMode (3x bounce/4x bounce activated)
INFORMATION
If only bouncing three times is activated in X-Soft, any number of bounces greater than or equal to 3 switches to
the first MyMode.
3 x
bouncing recognised
Switching by bouncing must be activated in X-Soft
4x bouncing must be activated in X-Soft
1. 3x bouncing on the forefoot
2. Remove load from prosthesis for 1 second
• Feedback from the prosthesis: 1 long vibration signal
3. Wait for 1 second
• Feedback from the prosthesis: 2 short beep signals
Switching from basic mode to the second MyMode
4 x*
bouncing recognised
* Switching by bouncing must be activated in X-Soft
* 4x bouncing must be activated in X-Soft
1. 4x bouncing on the forefoot
2. Remove load from prosthesis for 1 second
• Feedback from the prosthesis: 1 long vibration signal
3. Wait for 1 second
• Feedback from the prosthesis: 3 short beep signals
52 | Ottobock
52 | Ottobock
1 x
1 x
1s
for switching
with an extended leg and contact with ground throughout (within 1 second)
(no ground contact)
1 x
1 x
1s
for switching
with an extended leg and contact with ground throughout (within 1 second)
(no ground contact)
1 x
1 x
1s
for switching
with an extended leg and contact with ground throughout (within 1 second)
(no ground contact)
1 x
1 x
return to
basic mode*
and 1 long beep signal
and 1 short vibration signal
2 x
2 x
first MyMode
activated
and 1 long beep signal
and 2 short vibration signals
3 x
3 x
second
MyMode
activated
and 1 long beep signal
and 3 short vibration signals
.
.
.
.
.
.
Genium - Bionic Prosthetic System
Genium - Bionic Prosthetic System

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis