TROUBLESHOOTING
TROUBLESHOOTING
Equipment
See diagram 333 232 on page 74.
S
For column and boom versions see diagram 443 386 on page 78.
Universal meter
S
2 types of extension cards for Euro card format:
S
1.
2.
Check
that the control unit is correctly connected as shown in diagram
S
333 232,
see diagram 443 386 for column and boom versions
that cables and connections, including those for external equipment,
S
are undamaged and correctly connected.
that the controls are correctly set.
S
that all supply voltages are correct.
S
Note
that the 3 row connectors on the meter card are connected to the dis-
S
play card.
that cables or circuit cards must not be disconnected while power is on.
S
Wait 20 seconds after switching off before doing anything.
that the circuit card contact rows are marked as follows:
S
a = nearest circuit card
b = middle row
c = Furthest from circuit card
1 = the top 3(2) connections
32 = the bottom 3(2) connections
that the system zero is connected to the shunt/electrode and does not
S
carry earth voltage.
POSSIBLE FAULTS
1. Symptom Fuse A 23 on back panel blows when power is switched on.
Shorting in either motor regulator
Cause 1.1
Action
Measure resistance between a18 -- a30 on both motor regulators.
If it is less than 10 ohm, replace the regulator.
Cause 1.2
Shorting in a drive or brake transistor.
Action
Measure the resistance between the transistor casing and each leg individ-
ually. The resistance should be higher than 10 ohm.
dga7f1ea
Model DIN 41612 type C equipped for a--c
Model DIN 41612 type D equipped for a--c
It is possible to reverse the connections!
- - 14 - -