Herunterladen Inhalt Inhalt Diese Seite drucken

Purge Valve Installation - Agilent Technologies Turbo-V 1K-G Bedienungshandbuch

Turbomolekularpumpen
Vorschau ausblenden Andere Handbücher für Turbo-V 1K-G:
Inhaltsverzeichnis

Werbung

17
Technical Information

Purge Valve Installation

Purge Valve Installation

A gas purge valve is available to protect the pump bearings against
particulate and corrosive gases that could move into the pump.
To install the gas purge valve it is necessary to unscrew the purge
port cover as shown in the following figure.
Figure 38
and then screw the gas purge valve (with a torque of 2.5 Nm) as
shown in the following figure.
Figure 39
280/294
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)