17
Technical Information
RS 232/RS 485 Communication Description
When a data is indicated between two quotes ('...') it means that the indicated
NOTE
data is the corresponding ASCII character.
The field length is variable according to the data type as per the
following table:
260/294
<STX> (Start of transmission) = 0x02
<ADDR> (Unit address) = 0x80 (for RS 232)
<ADDR> (Unit address) = 0x80 + device number (0 to 31) (for RS
485)
<WIN> (Window) = a string of 3 numeric character indicating the
window number (from '000' to '999'); for the meaning of each
window see the relevant paragraph.
<COM> (Command) = 0x30 to read the window, 0x31 to write into
the window
<DATA> = an alphanumeric ASCII string with the data to be
written into the window. In case of a reading command this field
is not present.
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)