Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies Turbo-V 1K-G Bedienungshandbuch Seite 262

Turbomolekularpumpen
Vorschau ausblenden Andere Handbücher für Turbo-V 1K-G:
Inhaltsverzeichnis

Werbung

17
Technical Information
RS 232/RS 485 Communication Description
Tab. 6
Type
Logic
Numeric
Alphanumeric
ACK
NACK
Unknown
Window
Data Type Error
Out of Range
Win Disabled
262/294
Response types
Length
Value
1 byte
-
6 bytes
-
10 bytes
-
1 byte
(0x6)
1 byte
(0x15)
1 byte
(0x32)
1 byte
(0x33)
1 byte
0x34)
1 byte
(0x35)
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)
Description
After a read instruction of a logic
window
After a read instruction of a numeric
window
After a read instruction of an
alphanumeric window
The command execution has been
successfully completed
The command execu-tion has been
failed
The specified window in the
command is not a valid window
The data type specified in the
command (Logic, Numeric or
Alphanu-meric) is not accorded with
the specified Win-dow
The value expressed during a write
command is out of the range value
of the specified window
The specified window is Read Only
or temporarily disabled (for ex-ample
you can't write the Soft Start when
the Pump is running)

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis