Supplementary information
X-Porte Desktop
Assembling X-Porte Desktop
Caution: Before you assemble X-Porte Desktop, you must configure the input and output voltages
on the isolation transformer. Incorrect voltage settings can cause damage to the equipment. For
instructions, see the Toroid Superior Transformers Voltage Selections W series IsoBox, included with the
isolation transformer.
Troubleshooting
If you do not hear system sounds when you use the system on the desktop, make sure that you have
configured audio settings correctly (see "Audio Settings" in X-Porte Help). If you still do not hear
system sounds, cycle power: Turn the isolation transformer power switch to the OFF position and
then turn it to the ON position.
Turning on the system
If the system does not maintain expected battery charge, or if the battery icon on the clinical monitor
does not display the battery charge status, disconnect and reconnect the system to AC power.
Connect the system to AC power to maintain battery charge, especially if the system will not be used
for several days.
1 Turn off the system.
2 Disconnect the system from AC power.
3 Reconnect the system to AC power.
The battery charge indicator at the base of the stand blinks green, and the battery icon on the
clinical monitor displays the battery charging state.
Importing and exporting
You must have administrative privileges to import or export labels, exam types, user accounts, and
system settings.
3