Herunterladen Inhalt Inhalt Diese Seite drucken

After Use; Troubleshooting; Maintenance; Maintenance Intervals - Dräger Prestor 5000 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
3.6

After use

3.6.1
Work after testing
1. Log off from the system.
2. Disassemble the test setup.
3. Empty the nebuliser water container and allow to dry.
3.6.2
Disassemble the test device
1. Pull the latch of the table mount backwards.
2. Lift the test device at the back and pull out forwards from
the table mount.
4

Troubleshooting

Fault
Cause
The test device
Test device faulty
makes a hissing
sound when it is
switched on.
The LED on the
Test device faulty
test device does
not light up, even
though the test
device is
connected to
power supply and
a test has been
opened.
No communication
USB cable faulty
with the PC
software is
possible (the test
Test device faulty
device is not
recognised).
5

Maintenance

5.1

Maintenance intervals

Work to do
Test the gel face
1
Conduct inspection
Conduct a major overhaul
1
Only by trained specialist personnel.
2
To ensure the availability of test devices and thus support the
workflow in respiratory protection workshops, major overhauls on
test devices should be carried out as required by Dräger. Based on
the various uses for the test devices, test devices are subject to
different types of wear, but essentially Dräger recommends having
a major overhaul conducted at least every six years.
Dräger Prestor 5000
Remedy
Contact service
personnel
Contact service
personnel
Use a new cable
Contact service
personnel
X
X
2
5.2

Cleaning

If required, clean the test device with a moist leather cloth
and dry it.
5.3

Maintenance work

5.3.1
Replacing the gel face
1. Remove the old gel face from the moulded head section.
2. Moisten the inside of the new gel face with water.
3. To place the new gel face onto the moulded head section,
first position the eye with the measuring point on the head
section, and then position the other eye. Press the gel face
against the head section.
4. Ensure that the edge of the gel face is flush around the
entire head section.
5. Dräger recommends mounting a full face mask to the
moulded head section until the moisture between the head
section and the gel face has evaporated (e. g. over night)
and then commencing the test.
6

Transportation

Only transport the test device when it is switched off. The test
device may not be used during transportation in a vehicle.
7

Storage

Store the test device in a dry place. Avoid direct sunlight and
temperatures above 60 °C.
8

Disposal

This product is not permitted to be disposed of as
domestic waste. It is therefore marked with the
adjacent icon.
Dräger will accept the return of this product free of
charge. For information contact the national sales
organisations and Dräger.
X
Troubleshooting
13

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis