3. Automated wash
• Program the washer using the following parameters:
Phase
Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry Phase
• If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
Prolystica® 2x Neutral Detergent is validated for cleaning efficacy.
4
Low Level Disinfection (optional)
1.
Clean and prepare the camera head and cable as recommended in this
user guide. Ensure the soaking cap is installed.
2.
Disinfect the device in a disinfecting solution that has one of the
following active Ingredients:
• ≥ 2.4% glutaraldehyde
25°C
• ≥ 3.4% glutaraldehyde
25°C
• ≥ 0.55% ortho-phthalaldehyde
12 min at 25°C.
3.
Prepare the disinfecting solution according to the manufacturer's
instructions.
4.
Thoroughly rinse and flush all parts and lumens with running,
demineralized water to remove the disinfectant.
5.
Dry all parts with a lint-free towel immediately after rinsing.
CIDEX Activated® is validated for disinfection efficacy.
5
CIDEX Plus® is validated for disinfection efficacy.
6
CIDEX® OPA is validated for disinfection efficacy.
7
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
with a minimum soaking time of 45 min at
5
with a minimum soaking time of 20 min at
6
Water
Temperature
Cold tap
water
Hot tap water
Set point
(66 ˚C)
Hot tap water
115 ˚C
with a minimum soaking time of
7
Detergent Type and
Concentration
(if applicable)
N/A
Enzymatic Detergent
Non-enzymatic
Detergent
4
N/A
N/A
EN-31