SoftCloud™ Ace, SoftCloud™ Pro
Warning/Fault
Warning LED
+ audible warning
Warning LED
Patient is sinking or
"bottoming out" whilst
lying flat on the Mattress
If the problem is not resolved, please contact your Invacare provider.
30
Cause
Power failure
Service needed
The pressure may be set
too low for the patient's
weight
Solution
1. Press the "Audible Warning Mute" button to silence the audible
warning.
2. Check the power cable is firmly plugged into the mains outlet and
the Control Unit; and check the mains power is switched on.
3. Check the Control Unit fuse (1 amp slow blow fuse) – fuse can be
released using a screwdriver to push and turn.
If power is restored within 20 minutes of failure, the system
will run an initial start up sequence before returning to the
last settings.
1. Reset the warning with the "Audible Warning Mute" Button.
2. Contact an authorised Invacare representative.
There is no risk for continued use even if the indicator LED is lit.
1. Increase the pressure setting by pressing up the Pressure arrow.
2. To check effective system performance, conduct a "Bottoming Out"
test, see section "Using the mattress system".
1630732-A