SoftCloud™ Ace, SoftCloud™ Pro
Cleaning intervals
Regular cleaning and disinfection enhance smooth operation,
increase the service life and prevent contamination. We
recommend cleaning and disinfecting the product
•
regularly while it is in use,
•
before and after any service procedure,
•
when it has been in contact with any body fluids,
•
before using it for a new patient
6.2.2 Cleaning Instructions
IMPORTANT!
– The product does not tolerate cleaning in
automatic washing plants, with high-pressure
cleaning equipment or steam.
A cleaning record should be kept as part of cleaning
the system.
Cleaning control unit
1. Wipe down the control unit casing and hose fittings
with a damp cloth and suitable detergent.
2. Using a nylon brush, gently clean all crevices as they
can harbor microorganisms.
3. Wipe off the control unit casing and hose fittings with a
damp cloth to remove all detergent.
4. Air dry all treated surfaces.
Cleaning hangers
1. Wipe down the exterior of the hangers periodically,
using a damp cloth and suitable detergent.
24
Cleaning covers
(Removal of contaminants such as dust and organic matter)
1. Remove all covers for laundering.
2. Launder the covers with the recommended temperature
at 95 °C using a diluted detergent solution (Instructions
on label).
IMPORTANT!
Washing at higher temperatures will cause
shrinkage.
Drying covers
1. Hang covers from a line or bar and drip dry in a clean
indoor environment or tumble dry on a low heat setting.
IMPORTANT!
– Tumble dry setting must not exceed 40 °C.
– Do not tumble dry for longer than 10 minutes.
– Dry thoroughly before use.
6.2.3 Disinfection Instructions
IMPORTANT!
– Only use disinfectants and methods approved
by your local infection control institution and
follow your local infection control policy.
– Additional information on recommended
disinfectants can be found on
https://vah-online.de/en/for-users.