- Place the control unit in position on a fixed surface, horizontal or with a slope of no more than 5°.
- Check
that
the power switch is in position O (off).
- Connect
the
power supply cord to the receptacle of the device.
- Slot the plug (in compliance with the standards applicable in your country) onto the transformer and
connect it to the main.
- Connect the control footswitch to the footswitch connector.
Position
the footswitch so that it is easily accessible for your feet.
-
- Insert the bracket into the bracket holder on the top of the unit.
Connect
the micromotor and cord to the connector.
-
- Place the micromotor on its holder.
- Hang
the
bottle or bag of physiological saline from the bracket.
4.4 FIRST USE
Before using the I- S
URGE
sterilized according to the procedures defined in chapter X.
V – IRRIGATION SOLUTION
The I-
S
™ is not designed to administer drugs. It must be used only with bottles or bags of physiological saline
URGE
or sterile water with a capacity not exceeding one liter.
VI– IRRIGATION LINE
The I-
S
™ must be used exclusively with autoclavable or disposable sterile irrigation lines distributed by
URGE
®
SATELEC
.
VII– ADJUSTMENTS / SETTINGS / INTERFACE
7.1 START UP
First switch on unit screen (Fig. 5)
The first row shows:
o the Program in use P: 1 to 5
o the Ratio
o the motor sense (Forward "FWD" or Reverse "REV")
The second row shows the speed in use "SPEED" in rpm
The third row shows the "TORQUE" in Ncm
The fourth row shows if the motor is ON or OFF and the irrigation flow rate [0-1-2-3-4].
In every program Programs [1 to 5] are set the same parameters by the manufacturer:
Ratio: 20:1
Speed:
o "Implant Ster" micromotor: 2000 rpm
o "Implant" micromotor: 1500 rpm
Flow: 3
Torque: maxNcm
Motor
sense: FWD
™ for the first time, all the equipment must be given routine maintenance and/or
11