7. Unscrew and remove the tunneling tool handle.
8. Grasp the tip of the tool with one hand while holding the straw in place with the other hand. Pull the
tunneling tool shaft out through the straw.
9. Push the lead or extension proximal ends through the straw, then withdraw the straw.
CAUTION: Do not tunnel splitter.
Note: If using the 2x8 Splitter and performing a permanent trial, the splitter tails may be tunneled
to the exit site.
10. For Permanent Trials: If using the Infinion 1x16 lead, proceed to "Lead Connection to Splitter" on
page 9. Otherwise, proceed to "Connecting the Lead Extension" on page 14.
11. For Permanent IPG Implantation, if extensions are used, proceed to the instructions for "Connecting
the Lead Extension" on page 14.
12. For Permanent IPG Implantation, if splitters are used, proceed to the instructions for "Lead Connection
to Splitter" on page 9.
13. For permanent IPG implantation, refer to the IPG instructions in the appropriate DFU for your SCS
system as listed in your Reference Guide.
Note: The following Codman Disposable Catheter Passers may be used in place of the Boston
Scientific tunneling tool:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the IPG pocket to the
midline incision using the standard technique.

Connecting the Lead Extension

1. Wipe clean the proximal end of the lead, then insert the proximal end into the lead extension connector
or splitter until it stops and the retention ring (long ring) is under the setscrew.
Note: If there appears to be an obstruction when inserting the lead into the lead extension
connector, use the hex wrench to loosen (counterclockwise) the setscrew and/or gently
rotate the lead to help advance the proximal end.
2. Ensure that the lead is fully inserted before tightening the set screw to prevent lead damage.
Percutaneous Leads Directions for Use
90970887-16 Rev B
14 of 401