1744/1743
Users Manual
Min/Max Values
Logging detects the highest and lowest voltage RMS values and the highest
current RMS value during the test interval, using a minimum resolution of
10 ms.
The response time can be set in PQ Log to the following:
•
0.5 or 1 line power period
•
200 ms
•
1, 3, or 5 seconds.
U
30
Measuring Interval
Figure 14. Logging Min and Max Values
Max. Value
Min. Value
t
egb017.jpg