Guidance and manufacturer's declaration – Electromagnetic Emissions
(as request by regulation EN 60601-1-2:2001)
The surgical aspirator SUPER TOBI CARRELLATO is intended for use in the electromagnetic environment
specified below. The customers or the user of the surgical aspirator SUPER TOBI CARRELLATO should
Emissions Test
Irradiated / Conducted emissions
CISPR11
Irradiated / Conducted emissions
CISPR11
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration – Electromagnetic Emissions
(as request by regulation EN 60601-1-2:2001)
The surgical aspirator SUPER TOBI CARRELLATO is intended for use in the electromagnetic environment
specified below. The customers or the user of the surgical aspirator SUPER TOBI CARRELLATO should
Immunity Test
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
Electrical fast transient / burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
Magnetic field
IEC/EN 61000-4-8
Conducted Immunity
IEC/EN 61000-4-6
Irradiated Conducted
IEC/EN 61000-4-3
Note U
is the value of the power supply voltage
T
assure that it's used in such an environment.
Compliance
Group 1
Class [B]
Class [A]
Complies
assure that it's used in such an environment.
Compliance
± 6kV on contact
± 8kV in air
± 2kV power supply
± 1kV differential mode
5%U
for 0.5 cycle
T
40%U
for 05 cycle
T
70%U
for 25 cycle
T
<5%U
T
for 5 sec
3A/m
3Vrms 150kHz to 80MHz
(for appliances that aren't life -
supporting)
3V/m 80MHz to 2.5 GHz
(for appliances that aren't life -
equipment)
10
Electromagnetic environment -
guidance
The surgical aspirator
SUPER TOBI CARRELLATO only used
RF energy only for its internal
functioning. Therefore its RF emissions
are very low and are not cause
interference in proximity of any
Electronic appliances.
The surgical aspirator
SUPER TOBI CARRELLATO can be
used in all environments, including
domestic and those connected directly
to the public mains distribution that
supplies power to environments used
for domestic scopes.
Electromagnetic environments -
guidance
Floors should be wood, conceret or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a
typical commercial environment or
hospital
Mains power quality should be that of a
typical commercial environment or
hospital
Mains power quality should be that of a
typical commercial environment or
hospital If the user of the surgical
aspirator SUPER TOBI CARRELLATO
request that the appliance operates
continuously, the use of a continuity
unit
is recommended.
The power frequency magnetic field
should be measured in the intended
installation location to assure that it's
sufficiently low.
-
-