5.4.2 Output Functions

HYDACLAB
has 2 possible analogue outputs.
In the sub menu Output 2 you can configure the
analogue output in the output function.
The changes you perform here will also change the
analogue sequence.
In the analogue output the user can select the
parameters to be shown and set the analogue
output signal to a 0 - 10 V signal or 4 - 20 mA.
The measured values for the sequence are frozen at its
start-up in order to ensure that the measured values always
refer to the condition during the start of the sequence.
For the Arrhenius value and the operating hours, the output
signal is scaled to a maximum value which is set by the user,
see
. For all other parameters the measuring ranges
5.4.1.3
are default settings.
If no parameter has been selected, the analogue output is
switched off.
If only one parameter has been selected a pure analogue
signal is output (no sequence).
For each output, several parameters can be selected. In this
case the values are shown as sequences, see
2 seconds. In between each shown parameter there is a pause of 2 seconds. The sequence starts
with a 100% value representing the start value. The sequence is followed by a pause of 32 seconds.
This means 7 values plus the start value have a sequence length of 30 seconds. The sequence
starts every 62 seconds.
The output values of the oil condition represent:
Output value oil
condition
0%
No statement can be made
25%
OK
50%
Warning
75%
Quick oil change
100%
Alert