1.5.2
Name plate
The name plate is located on the transmitter housing.
Fig. 1
1 Transmitter model
2 Manufacturer of transmitter
3 Order code with SAP no.
4 Serial number
5 Approved power supply, current
communications protocol
6 Configured parameters
7 Permissible ambient temperature
Important
The temperature range on the name plate (7) refers only to the permissible ambient
temperature range for the transmitter and not to the measuring element used in the measuring
inset.
1.6
Transport safety information
Observe the following information:
•
Do not expose the device to moisture during transport. Pack the device accordingly.
•
Pack the device so that it is protected from vibration during transport, e.g. through air-
cushioned packaging.
1.7
Safety information for electrical installation
The electrical connections may only be performed by authorized specialist personnel according
to the electrical plans.
Comply with electrical connection information in the instruction. Otherwise, the electrical
protection class can be affected.
The secure separation of contact-dangerous electrical circuits is only ensured when the
connected devices fulfil the requirements of the DIN EN 61140 (VDE 0140 Part 1) (basic
requirements for secure separation).
For secure separation, run the supply lines separated from contact-dangerous electrical circuits
or additionally insulate them.
CI/TTF300-X1
1
2
Automation
Products GmbH
3
O-Code: TTF300-Y0B4/OPT
4
Ser.-No: 3452348673
5
U = +11...42 V, I = 4...20 mA, HART
S
a
CFG: 2 x TC; Type K; 0°C...300°C
6
T
= -40°C...+85°C
amb
7
TTF300
TTF300
Made in Germany
2006
8323455772
HW-Rev: 1.05
SW-Rev: 01.00.00
IP6X, NEMA 4X
www.abb.com/temperature
8 Internet address of manufacturer
9 Level of protection
10 Software version
11 Hardware version
12 CE mark (EC conformity)
13 Refer to product documentation
14 Year
15 Country
Safety
15
14
13
12
11
10
9
8
A00115