ASPINA DO M INSTALLATION – OPERATION – MAINTENANCE
• To separate, label, packing and providing for decontamination of contaminated parts by course of
national regulations
• Liquidate dental suction unit according to local effectives statutes.
Inside parts of unit can be contaminated with biological material by reason incorrect
using. Before clearing and waste disposal pass on special institution for
decontamination.

TROUBLESHOOTING

12. INFORMATION ON REPAIR SERVICE

Guarantee and post-guarantee repairs are ensured by manufacturer or organizations and repairers
denoted by the supplier.
Warning !
Manufacturer reserves the right to make changes on the appliance that however will have no impact
on substantial characteristics of the appliance.

13. SOLVING COMMON PROBLEMS

Disconnect the appliance from the network before making any interventions therein.
Works in connection with the removal of failures may only be carried out by a trained
stomer service technician.
In case of repair of parts of device which might be contaminated please follow bellow mentioned
instruction:
• Disconnect the appliance from mains
• To follow the rules of personal hygiene for works with contaminated material
• Clean device according, section 9
• To separate, label, packing and providing for decontamination of contaminated parts by course of
national regulations
• To perform the repair of damaged parts
Failure
Aspirator does not work Mains voltage absent
Interrupted supply of el. energy
Power-supply switch off
Interrupted power cord
Loose clamp of terminal board
Interrupted motor wiring, defective
thermal protection
Defective mains fuse of the equipment
Full separation vessel
High current consumption
Suction unit is/was awash by liquid
Aspirator is switching
Defective control automatics
Problem and possible reasons
Check mains voltage in the socket, activate a
protective element in the electric distribution
(fuse, circuit breaker)
Check the power-supply switch position ( „I"),
switch on the power-supply switch (network
indicator goes on)
Replace the defective part
Tighten the clamps
Replace suction aggregate
Check the state of equipment mains fuse,
replace defective fuse (fuses 2 x T 6.3 A are
situated in the upper, narrowed part of the
aspirator)
Empty the separation vessel content
Defective starting capacitor, replace the
capacitor
Suction unit pass on repair to service – keep all
rools and norms for work with contamination
material
Replace the defective part
- 16 -
Way of removal
NP-DO M-34_03-2019