Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
OBJ_DOKU-173544-002-1-mehrere.fm Seite 18 Freitag, 31. März 2006 10:35 10
Check
Check the product for visible damage and wear:
1.
Damage to the product, e.g. cracks, rough surface, splin-
tering, discoloration.
Damage to the product insulation and/or cable/connector
insulation, e.g. cracks and breaks.
Bending of the product or parts of the product. Do not
attempt to straighten bent instruments!
If damaged, do not use this product!

Packaging

The product must be protected against damage during steriliza-
tion.
Pack the product in disposable sterilization packaging (single
1.
or double packaging) made from paper/polyethylene and/or
in a sterilization container.
Sterilization
Only sterilize products that have been cleaned and disinfected.
Please follow the recommendations issued by the sterilizer manu-
facturer concerning loading, handling, and drying times.
ERBE Elektromedizin recommends steam sterilization with the
method described below. If other sterilization methods are used,
ERBE Elektromedizin shall bear no responsibility.
Steam
Fractionated vacuum method (with adequate product drying), at
least 20 min. at 121 °C or 7 min. at 134 to 138 °C.
18