Herunterladen Diese Seite drucken

Error Messages - Agilent Technologies 929-0032 Bedienungshandbuch

Tsp controller

Werbung

4
Instructions for Use

Error Messages

Error Messages

For a certain type of failure, the controller will self-diagnose the error
and the messages described in the following table will be displayed.
Tab. 2
Message
Power AC fail
Heater defective
Filament x
defective
Cartridge
defective
Fault:
overtemperature
External
interlock
74/127
Description
Incorrect mains power frequency.
The Mini Ti-Ball is exhaust. The
sublimation is turned off.
Filament X, where X can be 1, 2 or
3, interruption (= filament
exhaust) during a sublimation
cycle with source type TSP and
the "TSP RECOVER" parameter
set in "MANUAL".
The sublimation will turn off.
All TSP filaments are exhaust.
The sublimation is turned off.
Controller internal over-
temperature. The ambient
temperature is too high or an
external overload occurs: the
sublimation is turned off.
The interlock is not connected.
TSP Controller User Manual / 87-900-138-01
Repair Action
Switch OFF the unit then
check the mains power and
the correct setting of the
voltage selector.
Replace the Mini Ti-Ball.
Select manually another
filament not exhaust or set the
"TSP RECOVER" parameter to
"AUTOMATIC"; in this
configuration when a filament
becomes exhaust, the
software selects automatically
one of the remaining filaments
and the relevant LED will be
ON. The led related to the
filament detected defective
starts to blink.
Replace the TSP cartridge.
Check the load (pump and
cable) and eliminate the cause
of the overload.
Reduce the ambient
temperature.
Connect the interlock: the pin
3 on the I/O connector must
be shorted to one of the
Common pins (pin 6, 7, 8 or 9).
It is also possible to plug the
mating connector supplied
with the controller.

Werbung

loading