Type 8202/8222
Fig. 8:
Mounting the display module
Assembly
→
Check that there is a seal
on the cover and that it is
not damaged. Replace it
if necessary.
→
G r e a s e t h e s e a l i f
necessary, using a com-
ponent compatible with
the seal material.
→
Fully tighten by hand to
guarantee tightness.
7.3.
install the ph or redox probe in
the sensor holder of the 8202
(without fluid)
probe head
H
compression washer
seal
A
English
Following instructions
are valid for a Bürkert
probe.
If you use a probe
from another supplier,
respect the related
instructions.
→
Remove the protective plugs
→
Check that dimension H on
the probe is between 34
and 46mm. If necessary,
adjust the height of the com-
pression washer.
→
At first use, apply water or
soapy water on the "A" seal.
→
Insert the seal into the groove
on the holder.
→
Insert the probe with its seal
into the holder from above.
17