Herunterladen Inhalt Inhalt Diese Seite drucken

bürkert 8222 Schnellstartanleitung Seite 15

Ph- / redox- oder leitfähigkeits-messgerät
Vorschau ausblenden Andere Handbücher für 8222:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Type 8202/8222
Technicaldata
6.3.4.
Data of connectors and cables
number of fixed
type of connector
connectors
1 male M12 fixed
5-pin female M12 connector (not supplied).
connector
For the M12 connector with order code
917116, use a shielded cable:
• diameter: 3 to 6.5 mm
• wire cross section: max. 0.75 mm
1 male M12 fixed
5-pin female M12 connector (not sup-
connector and 1
plied) and 5-pin male M12 connector (not
female M12 fixed
supplied).
connector
For the M12 connector with order code
917116, use a shielded cable:
• diameter: 3 to 6.5 mm
• wire cross section: max. 0.75 mm
6.3.5.
sensor specifications of the 8222
conductivity sensor c=0.01
• Measurement range
• Type of fluid
2
2
• 0.05 µS/cm to 20 µS/cm
• ultra-pure water, pure water
conductivity sensor c=0.1
• Measurement range
• Type of fluid
conductivity sensor c=1
• Measurement range
• Type of fluid
6.3.6.
ph/redox probe of the 8202
The pH or redox probe must satisfies the following specifications:
• combined probe;
• length: 120 mm;
• with PG 13.5 head;
• with an S7/S8 fixed connector;
• without temperature probe.
The specifications of the probe can be found in the manual
of the probe used.
English
• 0.5 µS/cm to 200 µS/cm
• pure water, industrial
wastewater
• 5 µS/cm to 10 mS/cm
• industrial wastewater,
wastewater
15

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

8202

Inhaltsverzeichnis