Steam
Note for United States users: For all autoclave-compatible devices, Stryker
recommends using steam sterilization instead of liquid chemical sterilization.
• Steam sterilization is intended only for camera heads and couplers
marked
• Rapid cooling, or "quenching, " the coupler after autoclaving will
result in product damage.
• The water used in the autoclave process must meet standards for
clean steam per AAMI ST 79 Appendix M – Steam Quality
requirements.
Sterilizer Type
Minimum
Temperature
Minimum
Cycle Time
Product
Configuration
Drying Time
Ethylene Oxide (EO)
Preconditioning parameters
Temperature
Exposure
Concentration (100% EO)
Temperature
Time
Chamber Humidity
Aeration parameters
Aeration Time
Temperature
EN-34
.
AUTOCLAVE
"Flash"
Gravity
132 - 137°C
(270 - 279°F)
10 minutes
Unwrapped
"Flash"
Gravity
Prevacuum
132 - 137°C
132 - 137°C
(270 - 279°F)
(270 - 279°F)
3 minutes
10 minutes
Unwrapped
Double
wrapped
60 minutes
55 ± 2°C (131 ± 4°F)
725 mg/L
131 °F (55°C) ± 5 °F
1 hour
70% RH (50-80 %) ± 5%
12 hours
35 – 55°C (95° - 131°F)
Prevacuum
132 - 137°C
(270 - 279°F)
3 minutes
Double
wrapped
60 minutes