4.
Specialty Button: Selects different surgical settings
5.
P Button: Controls image (picture) capture via remotes 1 and 2
6.
Up and Down Arrow Buttons: Controls image light or zoom level
7.
W Button: Controls white balance and image light or zoom level
8.
Camera Connector: Connects to the 1188HD Camera Head
9.
DVI Out 1: Digital video output
10. DVI Out 2: Digital video output
11. Remote Out 1: Connects to a video accessory remote switch
12. Remote Out 2: Connects to a video accessory remote switch
13. Fiber Outputs (optical): DVI output for connection to Lucent
connector fibers
14. S-Video 1 Out: Analog video output
15. S-Video 2 Out: Analog video output
16. SFB Connectors: Enables FireWire connection with Stryker FireWire
devices. Provides connection for remote diagnoses and future software
upgrades.
17. AC Power Inlet: Connects to a separable power cord, which can be used
for mains isolation
18. Equipotential Ground Plug: Connects to a potential equalization
conductor. The resulting medical electrical system shall follow all
applicable IEC 60601-1 requirements.
19. Fuse Panel: Contains two 1.6A 250V (slow blow, high breaking capacity
1500A) fuses
20. SIDNE® Port: Connects to the SIDNE® Console to enable voice
operation and/or graphic tablet control
21. Composite Out: Analog video output
EN-9