Herunterladen Inhalt Inhalt Diese Seite drucken

Bruker S8 TIGER Originalbetriebsanleitung Seite 72

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Introductory User Manual
Tripline Switch:
The tripline switch makes sure that:
ƒ
the cover of the sample entry position is mounted correctly.
ƒ
the high voltage cable is properly attached to the X-ray tube.
ƒ
the X-ray tube is inserted into the tube mount.
K1/K2:
The relays K1 and K2 are activated by the microcontroller located on the "safety board" inside the S8
TIGER.
The switching contacts of relays K1 and K2 are inserted in both safety line 1 and safety line 2. If all
safety conditions are met, K1 and K2 will be activated and their switching contacts will close (i.e. en-
able) both safety lines.
In case of an error, K1 and K2 will be turned off and their contacts will interrupt both safety lines. The
X-ray generator will stop generating high voltage.
3-8
S8 TIGER X-ray Spectrometer
DOC-M80-ZXX031 V4 – 02.2010

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis