"BATT". If this happens, either recharge the battery or select a lower rate to complete the
feed cycle.
• The lifetime of the battery is optimally maintained by using the pump on a regular basis on
battery until it is fully discharged followed by a full recharge via the adapter on mains power.
• In case of battery failure, caused by, for example, an excessive temperature condition (which
may resolve itself with time) or by a complete battery failure, the E, F and battery symbol will
flash. Turn the pump "off" and contact your healthcare provider or refer the pump to service.

Connection to an external alarm system

The power connector at the side of the pump offers the option of connecting the pump via
the Flocare
®
Infinity
Nurse Call (corporate code 35752) to an external alarm system.
This external alarm system may be with open or closed contact.
• Follow the instructions for use of the Flocare
®
Infinity

ALARM FUNCTIONS AND SAFETY FEATURES

In case any of the problems listed on the problem solver chart occur, the pump delivers an audible
and visual alarm and stops working. The back light of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "START/STOP" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "START/STOP".
The problem solver chart on the next page gives a clear explanation of the alarms.

MAINTENANCE

Cleaning:
• Always unplug the pump prior to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor)
with warm soapy water, a 5% bleach solution in water, or a multipurpose disinfectant cleaner.
• The Flocare
®
Infinity
pump may be rinsed by holding under a stream of warm, clean water.
Do not submerge the pump!
• Always maintain the three rollers on the rotor in a clean state to ensure they spin freely.
• The adaptor normally does not require cleaning. When desired, a dry or slightly damp cloth
may be used to clean the outside surface of the adaptor. Make sure the adaptor is
disconnected from the wall outlet.
Safety notes:
• The manufacturer recommends an inspection of the pump at an authorised service centre
every 2 years.
• If any fault occurs during use, or if the pump is dropped, it should be checked by authorised
technical personnel prior to use.
UK
• Replace the Flocare
®
Infinity
pump set every 24 hours to maintain delivery accuracy and
prevent the growth of harmful bacteria. Dispose of Flocare
properly, as required by local law.
-52-

PROBLEM SOLVER CHART

Condition
NO SET
PUSH STRT
END OF DOSE
Nurse Call to connect the pump.
AIR
BATT
the pump
remains working
Battery, "E" and
"F" flashing
FILL SET
OCC IN
OCC OUT
OCC OUT
(repeated)
LOCK
®
Infinity
disposable sets
Cause
Correction
• The set is not fitted or wrongly
• Stop the alarm by pressing "START/STOP"
fitted in the pump.
• Insert the Flocare
®
Infinity
cassette into the pump as
indicated on the blister packaging and close the door.
• Restart the pump.
• The pressure sensor area is
• Clean the sensors, reinsert the cassette in the pump
dirty
and restart the pump.
• The pump has been untouched
• Stop the alarm and prolong the hold mode with
in hold mode for 3 minutes or
another 3 minutes by pressing "START/STOP" key.
more.
• Program the pump and start it by pressing the
"START/STOP".
• The pump administered the
• Turn the pump off by pressing the "ON/OFF" key
installed dose = volume.
and hold it during 2 seconds or
• Clear the memory of the total volume administered
(see paragraph: clearing the memory), reprogram a
new feeding schedule and start the pump by pressing
"START/STOP".
• The air sensor detected an
• Replace the empty feeding reservoir and continue
empty pump set.
feeding. If necessary prime the set.
• The air sensor area is dirty.
• Make sure the air sensor is clean.
• The feeding set is not inserted
• Make sure the feeding set is properly inserted
correctly.
in the pump.
• The feeding set is used
• Replace the feeding set by a new one.
for more than 24 hours.
• The battery capacity is too low.
• Connect the adaptor to the pump and mains and
The pump is not able to deliver
charge the pump for approx. 6 hours. During
highly viscous fluids at high flow
charging the pump can be used.
rates with the present charge
level of the battery.
• Battery failure.
• Turn the pump "off", contact your healthcare provider
or refer the pump to service.
• The pump is priming the set.
• Press the "FILL SET" key another time to stop the
pump and bring it back in the hold mode.
• The pump detected an upstream
• Stop the alarm by pressing the "START/STOP" key.
occlusion between the pump
• Remove the feeding set out of the pump and check
and the feeding bag.
the permeability by flushing the line.
• Re-insert the feeding set in the pump and restart.
• The pressure sensor area is
• Clean the sensors, reinsert the cassette in the pump
dirty.
and restart the pump.
• The pump detected a
• Stop the alarm by pressing the "START/STOP" key.
downstream occlusion between
• Check the permeability of the feeding tube by
the pump and the patient.
aspirating liquid via the medication port.
• The pressure sensor area is
• Clean the sensors, reinsert the cassette in the pump
dirty.
and restart the pump.
• Calibration not yet completed
• Stop the alarm by pressing the "START/STOP" key.
on the current feeding set
• Start the pump by pressing the "START/STOP" key
and let it run only briefly.
• Stop the pump by pressing the "START/STOP" key
insuring that there has been no occlusion out alarm.
• Remove the feeding set from the pump and reinsert
the feeding set in the pump.
• Restart the pump by pressing the "START/STOP" key.
• Only the current feeding
• The programming feature is blocked in the set up
schedule is allowed for this
mode of the pump. Ask your healthcare professional
patient. Another feeding pro-
to modify this setting.
gram is not allowed by your
healthcare professional.
UK
-53-