5.6 Analog Output Type

If "Pi" is set as control characteristic, the analog 0-10V output (see chapter 3.6)
can be configured as recorder or control output.
outP:
Output of actual pressure (recorder output)
out/S1:
Output of a continuous control signal f. switching output S1 (2-state-
control) or evacuation (3-state-control).
out/S2:
Output of a continuous control signal f. switching output S2 (2-state-
control) or venting (3-state-control).
To select the analog output signal switch the VD9 into configuration mode as
described in chapter 5, so that the display shows "unit".
S1
Press the menu key (several times) until the display shows the currently selected
output type (for control characteristic "Pi" only!).
By means of the "Up" and "Down" keys you can now toggle between
"outP", "out/S1" und "out/S2".
set
Confirm with "Set" key.
The display then shows the next configuration point. In order to return to meas-
urement mode quit all further configuration points by pressing the menu key.
vd9s2-de-130515