Chapter 2
Remove lateral cover right from arm B.
Remove plug link (JP2) for measurement of
phase current arm B.
Plug multimeter for current measurement
into jacks.
Switch on power supply.
Set right stepper driver PCB at run current
potentiometer to phase current of 750 mA.
(Tolerances +/- 5mA).
Switch power supply off.
Switch power supply back on and check
current – adjust, if necessary.
Switch power supply off again.
Remove measuring lines.
Reconnect plug link (JP2) for measurement of
phase current arm B.
Reconnect connecting cable (X24)
(master/slave).
Put housing panels back in place.
Service Manual / Leica ST5020
Seitenabdeckung rechts von Arm B entfernen
Steckbrücke (JP2) für Phasenstrommessung
Arm B entfernen
Multimeter zur Strommessung in die Buchsen
stecken
Spannungsversorgung einschalten
Den rechten Treiber am Runstrom- Potentiome-
ter den Phasenstrom auf 750 mA einstellen
(Toleranzen +/- 5mA)
Spannungsversorgung abschalten
Spannungsversorgung nochmals einschalten
und Strom kontrollieren, gegebenenfalls korri-
gieren
Spannungsversorgung ausschalten
Messleitungen entfernen
Steckbrücke (JP2) für Phasenstrommessung
Arm B wieder schließen
Verbindungskabel (X24) (Master/Slave) wieder
anschließen
Gehäuseabdeckungen montieren
Draft-Version 04/02
© Leica Microsystems
Mechanics
Page 120