7. Starting, operation
Tare Function DG-10-E
In this mode the positive deviation of the measured value from the tared value (max. 20% of
the measuring range) is constantly indicated by the main display. The untared pressure value
is always shown on the bar graph display. Thus the digital gauge's actual working load can be
read off even if the tare function is active. When the tare function is active, a function
indicator
appears on the display. There are 2 possible ways of activating this mode:
1) Select the tare function and set the value to be tared using the keys
When set, the measured value to be tared is backed up as tare. The main display and the
auxiliary display then show the value "0000" in the unpressurised state.
Example: If a value to be tared is set as 2 bar, the main display indicates 0 bar in the
unpressurised state.
2) Load the digital gauge with the pressure to be tared.
The current measured value appears in the main display field of the main display. Select
the tare function and then operate both keys simultaneously
measured value will be backed up as tare. The main display and the auxiliary display are
set to (0000).
Setting the tare value to (0000) resets the tare function in both models.
1
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio
GB
7. Starting, operation
Specifications
Pressure ranges
Over pressure safety
Burst pressure
Pressure ranges
Over pressure safety
Burst pressure
.
Pressure ranges
Over pressure safety
Burst pressure
Pressure ranges
Over pressure safety
Burst pressure
Materials
. The current
Wetted parts
»
Pressure connection
»
Pressure sensor
Case
Power supply
Operating time
Internal sampling rate
Insulation voltage
Display accuracy
Zero offset
Adjustability zero
Hysteresis
DG-10
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio DG-10
Model DG-10
bar
2
5
10
bar
5
10
20
bar
6
12
25
bar
160
250
400
bar
320
500
800
bar
1000
1200
1700
psi
30
60
145
psi
70
145
290
psi
85
170
360
psi
2000
3000
5000
psi
4640
7250
11600
psi
14500
17400
24650
1.4571
Ceramic Al
O
96%, NBR {EPDM}
2
3
XM-13 (1.4534)
1.4301
2x 1.5 V Mignon cell AA
h
4000 (AA 2000 mAh)
ms
200
VDC
500
≤ ± 0.5 ± 1 Digit
% of span
(according to IEC 61298-2)
≤ 0.1
% of span
(Power-up reset)
≤ 20
% of span
(via Tara-Function with model DG-10-E)
≤ 0.1
% of span
GB
20
50
100
40
100
200
50
120
800
600
1200
2400
300
600
1450
580
1450
2900
725
1740
11600
7500
10000
17400
21750
34800
43500
(up to 0 ... 50 bar)
(as of 0 ... 100 bar)
1