Connecting the OR Cable Assembly

The OR cable assembly is designed for temporary connection of a lead to the Trial Stimulator. A
cable extension is provided. When using a Splitter 2x8 prepare two OR Cable Assemblies.
CAUTION: Do not immerse the OR cable connector or plug in water or other liquids. The
OR Cable Assembly is intended for one-time only use; do not resterilize.
1. If two leads are being implanted, wrap the non-sterile "1-L" and "2-R" labels around the
cables at the Trial Stimulator to identify lead connections.
2. Verify that the Trial Stimulator is off.
CAUTION: Always turn the Trial Stimulator off before connecting or disconnecting the cable
assemblies.
3. Check that the locking lever on the OR cable connector is in the open position, marked "0".
4. For Percutaneous Leads, remove the steering cap from the stylet and slide the proximal
end of the lead with stylet, or splitter, into the open port on the OR cable connector.
5. Push the end of the lead or splitter into the port until it stops. Hold the lead in place while
sliding the lever to the locked position, marked "1".
6. Plug the OR Cable Assembly into the corresponding Trial Stimulator port(s) labeled "1-L"
(left) and "2-R" (right).
Clinician Manual
91173656-02
28 of 385
Production
Precision Clinician Manual N EU Entrada
Boston Scientific Confidential. Unauthorized use is prohibited.
locked
92070399 B.2