Diagram of climatic conditions:
% RH
100
90
80
70
60
50
40
30
20
10
0
-50 -40 -30 -20 -10 0
Altitude: < 2000 m
Use indoors or outdoors.
5.5. CONSTRUCTION SPECIFICATIONS
Overall dimensions of the unit (L x l x h): 270 x 250 x 180 mm
Weight: approximately 4.3 kg
5.6. COMPLIANCE WITH INTERNATIONAL STANDARDS
Electrical safety as per:
Double insulation
Pollution level: 2
Measurement category: III
Max. voltage relative to earth: 1000 V (2500 V in measurement category I)
5.6.1. ELECTROMAGNETIC COMPATIBILITY
IEC 61326-1 (Ed. 97) + A1, industrial environment category
5.6.2. MECHANICAL PROTECTION
IP 53 per IEC 60529 (Ed. 92)
IK 04 per IEC 50102 (Ed. 95)
3
2
1
10 20 30 40 50 60 70 80 90
1: Reference range
2: Operating range
3: Storage range (without battery)
IEC 61010-1 (Ed. 2 for 2001), IEC 61557 (Ed. 2005)
°C