LCD DISPLAY:
DESCRIPTION:
Remaining battery capacity has been
completely used. Battery requires charging.
Unit is turning off!
ACKNOWLEDGEMENT REQUIRED:
None required. Unit turns off within 30 seconds. Find alternate power and restart
normally.
LCD DISPLAY:
DESCRIPTION:
Unit is too hot, or too cold, to allow it to
continue to operate.
Unit will continue to run internal fans to
help lower, or raise, internal temperature.
Fans and unit will turn off after 10 minutes
regardless of temperature.
ACKNOWLEDGEMENT REQUIRED:
While the unit is warming / cooling the alarm screen will remain active. The unit will
turn completely off (battery mode) or return to standby screen (AC & DC mode),
whenever the desired temperature is reached or 10 minutes of cooling has expired.
Part No 1163190
SECTION 8—TROUBLESHOOTING GUIDE
INDICATORS:
Ten audible beeps every 10 sec
RED indicator Flashing Fast.
SOLUTIONS:
1. Connect transportable oxygen
concentrator to either AC or DC power,
or use an alternate charged battery
module.
2. If no other power source is available,
Change to another source of oxygen.
INDICATORS:
Triple audible beep every 10 sec
YELLOW indicator Flashing.
SOLUTIONS:
1. Move unit to warmer, or cooler,
surroundings. Allow unit to cool down to
less than 95° F (35° C), or warm up to 50°
F (10° C). Use AC, or DC, adapter for
power.
2. Clean intake filter.
3. Turn unit Off, then On again to retry.
4. Change to another source of oxygen
while waiting