Error codes

Any malfunctions during operation will be identified by an error message on the display, see e.g. fig. 29.
Gehen Sie dann wie folgt vor:
Disconnect the power supply for at least 10 sec.
Carry out corrective measures
Restart device
Error code
Er 1
Er 3
Er 5
Er 6
Er 7
Er 8
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
012007
Cause
Sensor not connected or defective
(Sensor rupture)
Temperature inside device higher than 76 °C
Sensor not in media
(no change in temperature of the
media after heater active for 3 minutes)
Temperature sensor short-circuit
The temperature of the medium is higher
than the maximum target temperature that
can be set
Calibration error
Effect
Correction
Device off
- Connect temperature sensor
- Replace temperature sensor
Device off
- Switch off the device and allow to cool down
Device off
- Insert the temperature sensor in the media to a depth of at least
20 mm.
- Increase the "Time-out" period
Device off
- Switch off the device and replace temperature sensor
Device off
- Switch off the device and allow the media to cool down
- Increase the maximum target temperature that can be set
Device off
- Switch off the device
- Re-calibration by authorised service personnel
23