Customer / med. Facility / practice:
Address:
Carried out:  repeat inspection
 _____________________________________________
Equipment type:
 hospital bed
Bed type:  DALI LOW ENTRY
Location:
Grade level (IEC60601-2-52):  1  2  3  4  5
Manufacturer: Burmeier GmbH & Co. KG
Testing equipment used (Type / Inventory Number):
MPG Classification: Class I
I. Visual Inspection
Visual inspection of the electriacl components
Stickers and type plates
housing of transformer unit / Power adapter
Motor housing and lifting rods of motors
Motor bolt
hand switch
Motor-, hand switch- and power cables
Plug and plug covering strip of control unit
Visual inspection of the mechanical components
Stickers and type plates
Safe working load / patient weight
Trapeze pole / socket
bed frame
Spring wood slats
castors
reclining surface
wooden boards
welded seams
Side rail
Knurled screws
wearing parts such as joints
locking ring of backrest/thigh rest
II. Electric measurments according to DIN EN 62353:2008-08 (leakage current, direct)
Proceed as follow:
1.
Plug the power cable into the test socket on the measuring instrument.
2.
connect the measuring instrument probe with an exposed conductive part of the bed frame (screws, etc.)
3.
activate the drive motors by using the hand switch for the duration of the measurement.
4.
Start procedures with the measuring instrument
result: bed SK ii (Type b)
G
P
ebraUchSaNweiSUNG
fleGebeTT
˛ Nursing bed
Presence, legible
Damage
correct position, Damage
check that the lock washer is properly in place
Damage, foil intact
Damage, routing
available, correct position
Presence, legible
Presence, legible
Damage, Deformation
Damage, Deformation
Damage, Splinters
Damage
Damage, Deformation
Damage, Splinters
Split welded seams
Damage, Splinters
Tight fit
Damage
check that the ring is properly in place
Value of first
limit Value
measurement
0,1 ma
ma
Dali lOw eNTrY · i
NSTrUcTiON
 inspection prior initial operation
 inspection following repairs / servicing
Protection classification:
Inventory Number:
Serial Number:
User-specific: none
1.
2.
Not
OK
Defect Description
OK
Not
OK
Defect Description
OK
current actual
value
ma
M
N
b
Dali lOw eNTrY
aNUal fOr
UrSiNG
eD
 i
˛ ii
Test protocol of
an inspection of
electro-medical
equipment
according to
DIN EN 62353:
2008-08
Page 1 of 2