3 GHz Spectrum Analyzer HM 5530
(Valid at 23 degr. C after a 30 min. warm-up period.)
Frequency characteristics
Frequency range:
150 kHz to 3 GHz
Frequency generation:
TXCO with DDS
(Digital Frequency Synthesis)
Stability:
± 5 ppm
Ageing:
± 1 ppm/year
Frequency resolution:
1 kHz (6 digit readout)
Center frequency range:
0 to 3 GHz
Tolerance of center frequency: ±2 kHz
Span setting range:
0 (zero span) and 1 to 3000 MHz
Amplitude characteristics
Display range:
–110 dBm to +20 dBm
Scaling, units:
10 or 5 dB/div, dBm, dBmV,
dBμV selectable
Dynamic range:
80 dB (10 dB/div), 40 dB (5 dB/div)
Amplitude frequency response (ATT 10 dB, zero span, 1 MHz-RBW
signal level –20 dBm):
±3 dB
Display (CRT):
8 cm x 10 cm
Display chracteristic:
logarithmic
Display units:
dB (dBm, dBmV, dBμV)
Input attenuator:
0 to 50 dB in 10 dB increments
Tolerance:
± 2 dB, referred to 10 dB
Maximum continuous input level:
Attenuation 0 to 50 dB:
+ 20 dBm (0.1 W)
Attenuation 0 dB:
+ 10 dBm
Maximum input dc voltage:
± 25 V
Reference level:
Adjustment range:
–110 dBm to +20 dBm
Tolerance (1500 MHz, ATT 10 dB, Zero Span,
RBW 1 MHz):
±1 dB
Min. average noise level (RBW 9 kHz):
150 kHz – 1.5 MHz:
–90 dBm
1.5 MHz – 2.6 GHz:
–100 dBm
2.6 GHz – 3.0 GHz:
–90 dBm
3
rd
order intermodulation (2 signals of):
–33 dBm each, frequency
difference › 3 MHz:
› 75 dBc
2
nd
order harmonic distortions (2
nd
ATT 0 dB, frequency difference › 3 MHz): › 75 dBc
Bandwidth dependent frequency response amplitude tolerance
amplitude tolerance (relative to RBW 1 MHz, zero span: ± 1 dB
Digitization:
± 1 Digit (0.4 dB) at 10 dB/div
scaling (average, zero span)
Marker/Deltamarker
Frequency resolution:
span/2000, max. 1 kHz, 6 Digit
Frequency accuracy:
± (1 kHz + tolerance of center frequency
+ 0.02% x span)
Amplitude resolution:
0.4 dB, 3 digit
Bandwidths
Resolution bandwidths (RBW) at –6 dB: 1 MHz, 120 kHz, 9 kHz
Videobandwidth (VBW):
50 kHz, 4 kHz
with automatic selection of sweep time:
40, 80, 160, 320 und 1000 ms
w w w . h a m e g . c o m
HM5530E/190506/ce · Subject to alterations · © HAMEG Instruments GmbH · ® Registered Trademark · DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49 (0) 6182 800 100 · www.hameg.com · info@hameg.com
1
2
harmonic at a signal level of -30 dBm,
1
2
1
2
S p e c i f i c a t i o n s
Inputs/Outputs
Measuring input:
N connector
Input impedance:
50 Ω
VSWR (ATT 10 dB):
typ. 1.5 : 1
Testsignal output:
N connector
Output impedance:
50 Ω
Frequency:
50 MHz ± 1 kHz
Level:
–10 to 0 dBm in 0.2 dB increments
Accuracy of level:
± 1 dB
Supply voltage for
field probes:
6 V
Audio output (PHONE):
3.5 mm jack
RS-232 interface:
9-pin. sub-D
External trigger input:
BNC connector
Digital signal:
Low level:
0 to +0.8 V
High level:
+2.5 V to +5.0 V
Functions
Keyboard input:
Center frequency, span, start frequency,
stop frequency, marker, deltamarker,
reference level, test signal level
Rotary encoder input:
Center frequency, span, start frequency,
stop frequency, marker, deltamarker,
reference level, test signal level intensity,
focus, trace rotation, volume
MAX HOLD:
Peak detection
AVG (average):
Averaging
Reference spectrum:
memory depth 2 k x 8 Bit
SAVE/RECALL:
Storage and recall of up to
10 instrument settings
AM demodulation:
for the PHONE output
REMOTE:
Display of remote/local control
Readout:
8 parameter display fields, display of
keyboard inputs
Miscellaneous
Display (CRT):
D 14-363GY, 8 cm x 10 cm internal graticule
Acceleration voltage:
approx. 2 kV
Trace rotation:
adjustable on front panel
Ambient temperature range:
+10 to + 40 °C
Storage temperature:
– 40 to + 70 °C
Power supply:
105 to 254 V
approx. 37 W CAT II
Safety class:
I (EN/IEC 61010-1) with protective earth
Dimensions (W x H x D):
285 x 125 x 380 mm Adjustable handle,
as a tilt-stand or for convenient carrying
Colour:
techno-brown
Weight:
approx. 6.5 kg
Accessories supplied: Line cord, manual, CD-ROM, adapter N male to BNC
female
Optional accessories:
HZ70 (27-0070-0000) Opto interface
HZ520 (17-0520-0000) plug-on antenna
HZ530 (27-0530-0100) field probe set for emi diagnosis
HZ560 (27-0560-0000) transient limiter
impedance converter 50 to 75 ohms.
, max. 100 mA (2.5 mm jack)
DC
, 50 to 60 Hz,
AC
Subject to change without notice
29