1.2 KERN SFB non-verifiable models:
Switch on instrument and press
Press
,
CHK" will appear.
Press
repeatedly until "P3 CAL" appears.
Press
and the first menu item "COUNT" will appear.
Press
repeatedly until "CAL" appears.
Press
. If necessary press
appears.
Press
. Ensure that there are no objects on the weigh-
ing plate.
Wait for stability display, then press
The currently set adjustment weight will be displayed.
Change by pressing the navigation keys (See manual chpt
2.1.1); select desired setting; currently enabled digit will be
flashing.
Acknowledge by
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during self test,
balance will return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting weight
the display will show an error message, repeat adjustment
process.