Electrical connection
48
The information in this section relates to the self‐con
tained operation of the T40B, without HBM system solu
tions.
The supply voltage is electrically isolated from signal out
puts and shunt signal inputs. Connect a separated
extra‐low voltage of 18 V to 30 V to pin 3 (+) and pin 2
(
) of plugs 1 or 3. We recommend that you use HBM
cable KAB 8/00-2/2/2 and the appropriate sockets (see
Accessories). The cable can be up to 50 m long for volt
ages ≥24 V, otherwise it can be up to 20 m long.
If the permissible cable length is exceeded, you can feed
the supply voltage in parallel over two connection cables
(plugs 1 and 3). This enables you to double the permissi
ble length. Alternatively, install an on‐site power supply.
Important
The instant you switch on, a current of up to 4 A may
flow and this may switch off power supplies with elec
tronic current limiters.