13.2.1 Active/passive operation of TTY interfaces

With the integration of wire jumpers, TTY interfaces can be set to active or passive operation.
active transmitters
Pins 1 and 2 bridged; Pin 3: ground, Pin 4: signal
output.

Table 10: TTY transmitter: active/passive operation

active receiver
Pins 5 and 8 bridged; Pin 6: signal input, Pin 7:
ground.

Table 11: TTY receiver: active/passive operation

13.3 BNC Bush X7

Pin
1
Inner conductor; (+) IRIG input
2
Outer conductor; (-) IRIG input

Table 12: IRIG Input bush

13.4 BNC Bush X8

Pin
1
Inner conductor; (+) IRIG output
2
Outer conductor; (-) IRIG output

Table 13: IRIG Output bushes

Order No. 2XV9450-1AR53-BA
TTY transmitter
passive transmitters
Pin 1: open; Pin 2: (+) signal output; Pin 3: open;
Pin 4: (-) signal output.
TTY receiver
passive receiver
Pin 8: (+) signal input, Pin 6: (-) signal output
Pins 5 and 7: open