Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance And Care - 3M ESPE Elipar FreeLight Betriebsanleitung

Kabelloses polymerisationsgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 10
• after 10 sec of exposure time have elapsed (1 alarm
signal), 2x after 20 sec, 3x after 30 sec.
• every time the "Exponential" mode is by-passed.
Two acoustical signals are emitted
• every time the power-down mode is terminated by
pressing any of the buttons.
A 2 sec-error signal is emitted, if
• the handpiece over-heats,
• the rechargeable battery is discharged to an extent
that reliable polymerization is no longer ensured.
A short 0.5 sec-error signal is emitted, if
• an attempt is made to set a different exposure time in
"Exponential" exposure mode. Only 40 sec of exposure
time are available in this mode.
Operating Errors
Error
Cause
The 10 sec-LED
The residual battery charge
within the dark blue
has dropped below 10%.
frame flashes and
a short signal is
emitted 5x when-
ever a button is
touched or the light
is turned off.
The ongoing exposure The battery lacks sufficient
is interrupted (light
charge.
off signal is emitted)
followed by a 2 sec-
error signal; the device
switches to power-
down mode and resists
further activation.
A 2 sec-error signal
The handpiece has become
is emitted upon
overheated in the course
pressing of the
of use.
Start button.
The handpiece can be used
again once it has cooled down.
Solution
Place the handpiece in
the charger and re-charge
the battery.
Place the handpiece in
the charger and re-charge
the battery.
Allow the handpiece to cool
down.
- The handpiece can be
used again, once
pressing the Start button
successfully activates
the light.
Error
Cause
The red LED of the
The battery is defective.
charger flashes.
The green LED of
Power outlet carries
the charger fails to
no voltage.
light up, even though
the power cable is
connected to a power
outlet.
Charger is defective.

Maintenance and Care

Replacement of the Battery
Never place the handpiece in the charger without the
battery inserted in the handpiece!
Use 3M ESPE batteries only! The use of other manu-
facturer batteries or non-rechargeable batteries/primary
batteries is a potential hazard and may damage the
device.
Rotate the lid at the lower end of the handpiece
counterclockwise until it hits the stop, and then
remove the lid.
Push a small spatula or similar aid between the
cylinders of the battery until the snap-lock releases
so that the battery protrudes slightly from the hand-
piece body.
Remove the battery from the handpiece.
Solution
Replace the battery.
Use a different power
outlet.
Have the charger repaired.
7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis