Herunterladen Inhalt Inhalt Diese Seite drucken

Errors; Maintenance And Care; Customer Information; Warranty - 3M ESPE RotoMix Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 6
SEITE 4 - 148 x 210 mm - 149/0/-/51+ F -

7. Errors

Error
Time display is off even though
the unit is switched on
Motor fails to run when the unit
is switched on
Capsule is thrown out of the holder
Error message "Lo" and error signal
are issued
NOTE: The RotoMix device may be repaired by 3M ESPE or 3M ESPE-authorised personnel only.

8. Maintenance and Care

The device requires no maintenance. Clean all compo-
nents with a soft cloth and, if necessary, a mild detergent.
Under no circumstances use solvents or scouring agents
as they will permanently damage the plastic.
Do not allow detergent to enter the mixing unit.
The following agents can be used for disinfection:
Pursept-A (Merz) and FD 322 (Dürr).
Pursept-A and FD 322 are not available in all countries.

Customer Information

No person is authorized to provide any information which
deviates from the information provided in this instruction
sheet.

Warranty

3M ESPE warrants this product will be free from
defects in material and manufacture. 3M ESPE MAKES
NO OTHER WARRANTIES INCLUDING ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR
A PARTICULAR PURPOSE. User is responsible for deter-
mining the suitability of the product for user's application.
If this product is defective within the warranty period,
your exclusive remedy and 3M ESPE's sole obligation
shall be repair or replacement of the 3M ESPE product.
4
SCHWARZ
- 03-0330 (sr)
Possible cause
Power failure
Cover of the unit is open
Motor is overheated
Capsule holder is contaminated
Capsule is too large or heavy
(refer to Technical Data section)
Cover was opened during
mixing
Motor or bearing are defective
Capsule is too heavy

Limitation of Liability

Except where prohibited by law, 3M ESPE will not be
liable for any loss or damage arising from this product,
whether direct, indirect, special, incidental or conse-
quential, regardless of the theory asserted, including war-
ranty, contract, negligence or strict liability.
Information valid as of 05/03
Remedy
Check both the power cable and main
voltage
Close protective cover
Allow the unit to cool
Blast capsule holder with compressed air
Use suitable capsules only
Press "Start/Stop" to quit
Press "Start/Stop" to quit. For a short
time, the speed (in revs per sec) at which
the device was turned off is shown in the
display.
Have the unit repaired.
Press "Start/Stop" to quit

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis