securely placed in the insertion device with
the needle pointing straight ahead prior to
insertion, pain or minor injury may occur. To
avoid this, please ensure that the following
steps are taken gently so the infusion set is
not loosened / removed by accident.
8
Connect tubing to a filled reservoir. Prime
infusion set. Ensure there are no air bubbles
in the reservoir or tubing before loading into
the pump.
Caution: When priming the infusion set,
hold the set with the needle pointing down
to ensure that medication does not come in
contact with the adhesive backing paper.
9
Pull up gently to remove the adhesive
backing paper.
10 Prepare the inserter: Place fingers on lined
indentations. Pull up the center of the inserter
until you hear a "click".
11
Carefully remove the needle guard by gently
twisting and then pulling it off.
Caution: Before positioning the infusion
set over the previously cleaned insertion
site, place the tubing in the slot on the side
of the inserter to ensure that the tubing is
not caught underneath the device during
insertion.
12
Insert Set:
Position device over the insertion
site. Simultaneously press the round
indentations on both sides of the insertion
device to insert the set.
13
Push gently on the center of the inserter to
secure the adhesive onto the skin.
14
remove the inserter and introducer needle by
grasping the center of the inserter and pulling
it gently back. massage the tape thoroughly
onto the skin. Fill the empty space in the
cannula with the following amount of insulin:
u-100 insulin:
6 mm: 0.3 units (0.003 ml)
9 mm: 0.5 units (0.005 ml)
15
Push the lid back in place until you hear a
"click."
To Disconnect
A
This set allows you to temporarily disconnect
from your pump without having to change
the infusion set. hold the adhesive in place
by placing a finger just in front of the cannula
housing, and gently squeeze the sides of the
connector. Pull the connector needle straight
out from the cannula housing.
B
Insert the disconnect cover into the cannula
housing until you hear a "click".
E N G L I S H
To Reconnect
C
Prime the infusion set until medication
dispenses from the needle.
Caution: When priming the infusion set,
hold the needle pointing down.
D
Place a finger on the adhesive tape just in
front of the cannula housing.
E
Push the connector needle straight in until
you hear a "click".
11
11