Listing of calibration steps
(*) Wait for complete stabilization of the display.
Calibration of frequency compensation
While observing the necessary safety precautions
open the instrument. Select the 50 V AC range.
Apply 25 V AC 15 kHz. Adjust the capacitor CV1
until the display reads 25,000 ±5 digits.

WDM8012 SOFTWARE

A CD-ROM is provided with the HM8012 multimeter,
including the user manual, the WDM8012 software
running under Windows     , as well as an application
software running under Excel     .
34
All instrument functions can be controlled by a host
computer through serial interface available in
standard.
The WDM8012 software is a virtual instrument
allowing the command of the instrument and the
reading of its configuration. This one can be saved
and recalled later.
When the configuration of the instrument is done, a
set of measurements can be carried out and saved
for future use.
Furthermore, the software can show deviations of
the values relative to two predetermined thresholds.
A DDE link is also possible, which make easier the
integration of the instrument in specific application
software.
A software running under Excel     allows automatic
drawings of curves, with a programmable delay
between successive measurement ranging from 1s
to 65 s.