ADDITIONAL BALANCE FUNCTIONS:

By using a special holder, it is possible to carry out weighings below the weighing
chamber through the GD hole. The holder is not available from METTLER TOLEDO and
must be specially made by the customer.
GD hanger
– Open all sliding windows.
– Remove weighing pan.
– Place balance on its back.
– Loosen screw (14) on the bottom of the
balance.
– Swing out cover (15).
– Retighten screw (14).
In the opening is a hook (16). A hanger is placed
on this hook, and the object to be weighed is placed
on the hanger.
– Set balance back on its feet, place pan back on
and level balance.
– With the hanger attached, press tare.
The weighing pan does not have to be put back on
as long as the holder is at least as heavy as the
weighing pan.
10
Below the balance weighing and
remote control.
Remote control
Taring can be done by using an external hand key or
foot pedal (see Accessories).
– Connect hand key or foot pedal (17) to the
sockets (8) on the rear of the balance.

CARE AND MAINTENANCE

The balance requires very little care and maintenance. A cloth and some
soapy water are sufficient for cleaning the weighing pan and balance
housing. Do not use any strong solvents. Use the small brush (see
Accessories) to remove any foreign material from the weighing chamber.
Do not blow air into the chamber under any circumstances!
Replacing the microfuse
– Disconnect power cable.
– Remove cap (6) with a screwdriver.
– Replace defective fuse (replacement fuse is
located in holder).
– Put a new replacement fuse in the holder.
– Place fuse holder back in balance and lightly
press in.
– Reconnect power cable.

Quicklinks ausblenden: