OPERATION: How to set the integration time and the stability detector.
The AE163 is capable of providing excellent results in spite of unfavorable weighing conditions. The stability detector (ASd: automatic stability detection) provides an
added measure of reliability to the results.
Setting criteria:
Standard setting
Int : Step 2
ASd : Step 2
If the integration time setting or stability detector setting is changed, any previously displayed weighing result will remain unchanged.
Int
- 1 -
Integration time (Int)
– Press single control bar (1) until balance
displays Int.
– Choice can be made between Steps 1, 2 and 3
by briefly pressing the control bar.
Step 1: shortest time
Once the desired step has been selected, wait until
the balance is again in the weighing mode.
If the standard setting is too slow, select Step 1.
If the display is too unstable, select Step 3.
8
Increased weighing cycle speed
Int : Step 1
ASd : Step 2
ASd
- 1 -
Stability detector (ASd)
– Press single control bar (1) until the balance
displays ASd.
– Choice can be made between Steps 1, 2 or
OFF by briefly pressing the single control bar.
Step 1: greatest sensitivity
If use of the stability detector is not wished, it can be
turned off ("OFF" position), whereby DeltaDisplay is
also switched off.
Increased weighing accuracy
Int : Step 3
ASd : Step 1
The green pilot light (12) in the left of the display
goes off when the balance has achieved stability.
When the pilot lamp is on, the data interface is
blocked.
The result should always be read only after the pilot
lamp goes out.