Dependence On Gas Type
The output signal depends on composition and type of the gas being measured.
The unit is adjusted for N
be corrected below 0.1 mbar by setting correction factors for both sensor types
via RS485 (see chapter 4.2 and 4.4).
4.2

Communication

4.2.1

The Serial Interface Of The VSH82

The VSH82 transducer is equipped with a serial RS485 interface. To enable
communication via RS485 please remove the rubber cap over the address
switch (1) and then set the address switch to a value between 1 and 15 using a
small screw driver or a similar tool (2). Afterwards insert the rubber cap again.
Communication is carried out according to the Thyracont protocol, detailed
information hereto is provided in a separate description.
and dry air. For other gases the pressure display can
2
The analog output signal 0-10V is simultaneously available
when RS485 is activated by address setting 1-15!
35