Herunterladen Inhalt Inhalt Diese Seite drucken

ADM Messtechnik OCM160 Betriebsanleitung Seite 61

Dc - prozesskalibrator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

OCM160 DC-Prozesskalibrator
Description:
This command allows reading and changing sequence name. Data will be saved to non-
volatile memory on STEP:PRES:SAVE command.
Parameters:
<SPD>
Remarks:
Overlapped command
Example:
STEP:PRES:NAME "STEP 1s"
STEP:PRES:NAME? Response: "STEP 1s"
[:SOURce]:STEP:PRESet:PCLear
Syntax:
[:SOURce]:STEP:PRESet:PCLear
Description:
This command clears existing sequence data including its timing table. Data will be
saved to non-volatile memory on STEP:PRES:SAVE command.
Parameters:
None
Remarks:
Overlapped command
Example:
STEP:PRES:PCL
[:SOURce]:STEP:PRESet:RAPPend
Syntax:
[:SOURce]:STEP:PRESet:RAPPend <SPD>
Description:
This command appends new record at the end of timing table. Data will be saved to non-
volatile memory on STEP:PRES:SAVE command.
Parameters:
<SPD>
Remarks:
Overlapped command
Example:
STEP:PRES:RAPP "0.5,220.0"
[:SOURce]:STEP:PRESet:RCOunt?
Syntax:
[:SOURce]:STEP:PRESet:RCOunt?
Description:
This commands returns actual number of records in timing table.
Parameters:
<DNPD>
Remarks:
Overlapped command
Example:
STEP:PRES:RCO? Response: 6
[:SOURce]:STEP:PRESet:ROW<IND_ROW>:AMPLitude
Syntax:
[:SOURce]:STEP:PRESet:ROW<IND_ROW>:AMPLitude <SPD>
61
Quoted sequence name. Upper alpha, lower alpha, digits and spaces
are allowed. Maximum string size is 8 characters.
Quoted string representing amplitude. The amplitude consists of two
float numeric fields separated by comma. The first one represents
timing interval in seconds and the second one amplitude.
Integer value representing number of records.
Orbit Controls AG
Owner's Manual

Werbung

Inhaltsverzeichnis
loading