The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
STEP ON
STEP? Response: 1
[:SOURce]:STEP:SELect<IND_STEP>
Syntax:
[:SOURce]:STEP:SELect<IND_STEP> <DNPD>
[:SOURce]:STEP:SELect<IND_STEP>?
Description:
This command selects timing sequence. Selected sequence is shown on device display,
can be executed using OUTP ON command and can be edited using SOUR:STEP:PRES
group of commands. Sequence is defined by function and by number of timing sequence
for this function.
Parameters:
<IND_STEP>
<DNPD>
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
STEP:SEL1 0
STEP:SEL1? Response: 0
[:SOURce]:STEP:PRESet:COUNt?
Syntax:
[:SOURce]:STEP:PRESet:COUNt?
Description:
This command retrieves maximum number of timing sequences. This number represents
maximum index used in sequence commands.
Parameters:
<DNPD>
Remarks:
Overlapped command
Example:
STEP:PRES:COUN? Response: 64
[:SOURce]:STEP:PRESet:NAME
Syntax:
[:SOURce]:STEP:PRESet:NAME <SPD>
[:SOURce]:STEP:PRESet:NAME?
Betriebsanleitung
Range 1 ... 6 (1 - if omitted):
1 – Voltage function
2 – Current function
3 – TC function
4 – Frequency function
5 – RTD function
6 - Resistance function
Range 1 ... maximum sequence count (32), one based index of
sequence
Integer value representing maximum sequence count.
OCM160 DC-Prozesskalibrator
60