Operation / set-up
!
"
!
"
84
V
W
2.
Use
or
to scroll through the values, symbols and picklists.
3.
If the parameter is set correctly, press
Note!
• Any settings that are displayed in grey cannot be selected / cannot be changed (only
notes, or option not available/not activated).
• With the factory setting "0000" (delivery status), configuration is possible at any time.
It can be protected against unauthorised manipulation by entering a 4-digit release
code. This has to be entered when the settings are later changed, if the device
settings are to be changed by keyboard.
Tip: make a note of your release code. Store it somewhere where unauthorised
persons do not have access to it.
• The changed settings do not become effective until you return to normal operation by
Z
repeatedly pressing
with the previous data.
5.5
Set-up using PC
You can also put your Digital Recorder into service and/or configure it via PC. The RS
232 or RS 485 system interfaces at the rear (depending on scope of delivery) are
available for this purpose
Installation of PC software:
1.
Install the supplied PC software on your computer. The program's Operating
Instructions can be printed out after installation, if required.
2.
Call up the program.
"
Caution!
Please make the plug connection first, then the connection to your PC. When
removing the connection, proceed in reverse order.
3.
Now you can configure your device via PC. For this, please refer to the program's
instructions/help.
Advantages of configuration via PC:
• The device data is saved in a database and can be accessed again at any time.
• Text entries can be carried out more quickly and efficiently by keyboard
• Measured values can also be read out, archived and displayed on the PC with the
program.
Caution!
Configuration can only be carried out via one interface (RS 232 or Ethernet) at a time.
5.6
Set-up using diskette
Save the device settings on the PC onto the diskette using the operating and readout
software. This set-up file can be adopted into the device if released under "Diskette
functions - Load set-up from diskette".
Note!
In order to be able to use this function, the device must already be registered in the PC
database.
Caution!
If the set-up diskette is not removed, the measurement data will begin to be saved after
approx. 5 minutes. The set-up data is still kept. Please change the diskette if the
measurement data is not meant to be saved onto this diskette.
F
again to confirm.
F
and confirming with
. Until this time, the device still works
Digital Recorder