Software update or upgrade via PC software /
diskette
Caution!
When the program is updated, all measurement data in the memory and on the diskette
is deleted. If the measured values saved in the device are still required, they should first
be read out or the diskette should be updated and removed from the device. After the
program transfer, all device settings are reset to the factory settings.
1.
Install and start supplied PC software.
2.
Connect the device to the PC.
3.
Remove, if necessary, the write protection on the program diskette (only for
software upgrade).
4.
Insert program diskette into the PC disk drive
5.
Go to the "Miscellaneous / Special device functions / Digital Recorder / Transfer
program" menu.
6.
Select interface parameters (Com-Port)
7.
Select desired program file and confirm with OK.
5.10
Software update or upgrade using program
diskette at device
1.
Save the device set-up to a new diskette ("Diskette functions / Save set-up on
diskette")
2.
Switch off the device.
3.
Insert the new original program diskette into the device.
4.
When switching on again, keep the
flashes.
5.
Release the keys; the new program is loaded (takes approx. 5 minutes). Screen
remains dark during loading, green LED flashes. Confirm the reset message with
OK.
6.
Save the previously saved set-up parameters if required back into the device
("Diskette functions / Load set-up from diskette").
Caution!
When the program is updated, all settings in the memory and on the diskette are
deleted.
5.11
Communication via serial interfaces / modem /
Ethernet
5.11.1
RS232 / RS485
The RS 232 serial interface can be accessed from the rear (9-pin Sub-D socket).
Note!
It is not possible to use the RS232/ RS485 and the Ethernet interface simultaneously.
The desired interface type must be selected in the set-up under "Miscellaneous - Serial
Interface".
Alternatively to the RS232 interface, the RS485 interface can be accessed from the rear
of the device. The RS485 interface can be used for remote configuration/readout (up to
approx. 1000 m lead length).