4. Operation
Turning the furnace on
A
Connect the power cord with the wall socket.
Put the On/Off switch to position "I".
After the self-test, the display shows the Home screen. All functions of the Programat can
be selected from this screen. Click the HOME key in order to return to the Home screen.
Selecting the program type
B
The furnace comprises multiple program types
a. Programs for Ivoclar Vivadent materials
b. 700 individually adjustable programs
Selecting the program group
C
The program types are divided into program groups.
Each program group comprises 20 programs.
Selecting the program
D
Select the required program.
Loading the furnace
Open the furnace head by pressing the OPEN FURNACE HEAD key and
place the object to be fi red with the fi ring tray in the furnace.
Programat Infrared Technology: Please read Chapter 6.3 of the Operating
Instructions carefully before using the furnace for the fi rst time. The notes
regarding the Programat Infrared Technology must be observed!

Starting the program

E
Press the START key to start the program. The LED lights up green. You can see the progress in the fi ring curve. At the end of the program,
the furnace head opens automatically. As soon as the OSD lights up green, the furnace is available for the next program start.
A
B
D
C
E