Troubleshooting

Fault
» display « (Printer Text)
No function after power-on
» ZERO TEST INCORRECT «
(BLANK CHECK INCORRECT)
» MOUTH ALCOHOL «
(MOUTH ALCOHOL)
» MINIMUM VOLUME NOT ACHIEVED «
(MINIMUM VOLUME NOT ACHIEVED)
» BLOWINGTIME TOO SHORT «
(BLOWINGTIME TOO SHORT)
» NOT ALLOWED «
(NOT ALLOWED)
» READINESS TO BLOW EXPIRED «
(READINESS TO BLOW EXP)
» ALCOHOL IN THE ENVIRONMENT «
(ALCOHOL IN ENVIRONMENT)
» CHECK AIRWAY «
» DATA INPUT TIMEOUT «
(DATA INPUT TIMEOUT)
» INTERFERENCE «
(INTERFERENCE)
The following errors are indicated on the
screen in English language with an error
code and a text message:
» ERROR [XXX] ..
..MAIN-SYSTEM «
..IR-SYSTEM «
..EC-SYSTEM «
..FLOW-SYSTEM «
..PRESSURE-SYS «
..HEATER SYSTEM «
Example:
ERROR [118] HEATER SYSTEM
The error code and the text message are
printed on the printer.
Example:
Error [118] HEATER SYSTEM
Messages generated by the Windows CE operating system
(Usually full text messages are generated, in parts with options allowing to select further actions).
Fault / Display
Instrument not ready for operation.
Service function is not possible.
Dräger Alcotest 9510 Standard IR / EC
Cause
Fuses defective.
Power supply insufficient.
Zero adjustment incorrect.
Residual alcohol in the mouth of the
person being tested. Person has drunk
alcohol directly before test or has
burped.
The supplied volume is smaller than the
minimum volume required.
The time for which the breath sample is
delivered is shorter than the minimum
requested.
Breath sample delivery at a not permitted
time, instrument not ready for
measurement (reblow).
Readiness for measurement of 3
minutes has expired. Breath sample
delivery has not started within this time
window.
Contamination of ambient air detected.
Necessary volume for purging the
measuring system not reached.
No data input within 10 minutes despite
prompt.
The two independent measuring
systems have detected an impermissible
influence on the measuring values by
foreign substances in the breath.
Instrument defective
(XXX = code specifying the error))
Instrument blocks automatically all
measurement cycles when detecting this
error.
The set temperature for the sensor or
hose is not reached.
Cause
Selected action is only possible when
instrument is » READY «.
Troubleshooting
Remedy
Check the fuses, replace if required, see
page 26.
Check power supply.
Ensure that ambient air is clean and free
of alcohol. Air the test room.
Remedy cause. Repeat measurement
after waiting time.
Observe control time.
The instrument automatically requests to
repeat the breath sample.
The instrument automatically requests to
repeat the breath sample.
Instruct person being tested for correct
delivery of breath sample. Start new
measurement.
Start new measurement.
Ensure that ambient air is clean and free
of alcohol. Air the test room.
Check breathing hose inlet. If no
abnormalities are found: restart
measurement.
If fault occurs again: inform Dräger
Safety.
Start new measurement.
Enter data when prompted to do so.
Identify cause of the influence, remedy if
possible.
If required initiate other evidential
measures.
Switch off instrument.
Check whether the error re-occurs after
switching on again.
If yes: inform Dräger Safety. Submit
instrument for repair.
Ensure compliance with the ambient
conditions. Switch off instrument, if
required put on instrument cover and
switch on again. If the error should
reoccur upon switch-on, inform Dräger
Safety.
Remedy
Wait until readiness for operation is
established.
25